PIONEER

lijntekening_fit_track

Obvious solutions are not always the easiest to achieve

A next step in cardiovascular treatment should consist of increasing treatment effectiveness
while reducing risk and medication requirement. This has led to the obvious solution of
a CoCr stent on a drug eluting balloon.

Many innovations and many scientific studies have been required to achieve this reliable,
safe and effective great new opportunity for better care.

Pioneer Technical Data

Stent
Stent Material: F562 Cobalt Chronium
Metal to artery ratio: ~14%
Strut thickness:

0.0037" (0.095 mm)

Strut width:

0.0037" (0.095 mm)

Maximum expansion diameter:

4.7 mm

Average recoil: Low <4%
Average foreshortening: Minimal <2.5%
Crush resistance: High; 0.15N/ mm
Radiopacity High
Stent flexibility Very High

Delivery system
Nominal pressure: 11 bar
Rated Burst Pressure: 18 bar (except Ø 4.0 = 15 bar)
Balloon folds 3-folds; Wing-Seal Technology
Balloon material: Semi-compliant polyamide for controlled dimensions
Drug: Paclitaxel (3.0 µg/mm², Drug loaded balloon surface)
Guiding catheter compatibility: 5F (min ID 0.056”)
Guide wire compatibility: Max 0.014”
Catheter type: Rapid Exchange
Proximal shaft: 1.9F (0.64 mm)
Tip Short, thin and soft 3 mm tip
Distal shaft: 2.4F (0.80 mm)
Average crossing profile

1.0 - 1.1 mm (0.039" - 0.043")

Lesion entry profile: 0.016” (0.40 mm)
Catheter coating:
Hydrophilic coating


 

More information about the Pioneer

downloadbutton  download Brochure Pioneer - pdf (open in acrobat)

downloadbutton  download Brochure DEB Technology - pdf (open in acrobat)

downloadbutton  download Compliance Chart Pioneer - pdf (open in acrobat)


Pioneer is registered trademark of BLUE MEDICAL.

 

Indications, contraindications, warnings and instruction

for use can be found in the product labeling.
CAUTION: restricted to sale by or on the order of a physician.

Not for sale in the US.